Sign In | Join Free | My disqueenfrance.com
disqueenfrance.com
Products
Search by Category
Wholesale Marketplace
Home > Multimeters >

Bench Type LCR Digital Meters Inductance Measuring Instrument

Categories LCR Digital Meters
Brand Name: VICTOR
Model Number: 4090B
Certification: CE
Place of Origin: CHINA
MOQ: 1pcs
Price: 4090C: USD177/pc;
Payment Terms: T/T, Western Union, L/C, Paypal
Supply Ability: 1000pcs available
Delivery Time: 5 to 30 days
Packaging Details: 6pcs/Carton; Carton Size: 103*45*45cm;20.5KG/Carton
LCD Display: 3.5 inch TFT Display
Size: 265mmx105mmx305mm
Material: Plastic, metal, silica gel
Product Name: Digital LCR Meter
Model No.: 4090A
Packing Details: 6pcs/Carton; Carton Size: 103*45*45cm;20.5KG/Carton
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

Bench Type LCR Digital Meters Inductance Measuring Instrument

Bench Type LCR Digital Meters Inductance Measuring Instrument


VICTOR 4090C Bench Type Digital LCR meter Capacitance Resistance Impedance Inductance Measure Instrument Electric Bridge Component​

1.1 General Technical Specifications

●Power supply voltage: 220V AC ± 10%, or 110V AC ± 10%, 45-65Hz
●Working environment: 0 ~ 40 ℃, relative humidity <80%.
●Storage environment: -10 ~ 50 ℃, relative humidity <80%.
●Dimensions: 265mm * 105mm * 305mm (width * height* depth).
●Mass: 2.3kg.
●Power consumption: <10W

SNDescription
1Power jack: AC 220V / 50Hz power supply input socket.
2Voltage selector: 110V / 60Hz VAC or 220V / 50Hz VAC.
3RS232 INTERFACE.
4USB Device interface.

The following data shall be measured under the following conditions:

  • Temperature condition: 23℃±5℃
  • Humidity condition: ≤65% R.H.
  • Zero value adjustment : open circuit and short circuit zeroed before the test
  • Warm-up time:> 30 minutes
  • Calibration time: 12 months

C: 0.20%*(1+ Cx/Cmax+ Cmin/Cx )(1+Dx)(1+ks+kv+kf);
L: 0.20%*(1+ Lx/Lmax+ Lmin/Lx )(1+1/Qx)(1+ks+kv+kf);
Z: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(1+ks+kv+kf);
R: 0.20%*(1+ Rx/Rmax+ Rmin/Rx )(1+Qx)(1+ks+kv+kf);
D: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(1+Dx+Dx*Dx)(1+ks+kv+kf );
Q: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(Qx+1/Qx )(1+ks+kv+kf );
The maximum and minimum values of the measurement parameters that affect accuracy

Frequency(Hz)1001202004008001K2K4K8K10K
Cmax800667400200100804020108
Cmin150012507503751871507537.518.715
Lmax15901325795397.5198.815979.539.819.915.9
Lmin3.22.61.60.80.40.320.160..080.040.032
Z/Rmax1
Z/Rmin1.59

Frequency (Hz)15K20K40K50K80K100K
Cmax5.3421.610.8
Cmin107.53.7531.871.5
Lmax10.67.953.983.181.991.59
Lmin0.0210.0160.0080.00640.0040.0032
Z/Rmax1
Z/Rmin1.59

Wherein: the unit of Cmax is uF; the unit of Cmin is pF; the unit of Lmax is H; the unit of Lmin is mH; the unit of Zmax / Rmax is MΩ; the unit of Zmin / Rmin is in Ω;

Technical Data

Model4090A Type4090B Type4090C Type
DigitsMain parameter: 5 digits; secondary parameter: 5 digits
Measurement ParameterMain parameter: L/C/R/Z; secondary parameter: X/D/Q/θ/ESR
Measurement Range(L)0.01μH - 9999H
Measurement Range(C)0.01pF - 99999μF
Measurement Range(R)0.0001Ω - 99.99MΩ
Basic Accuracy0.2%
Test Frequency (Hz)

10 point(100,120,200,400,
800,1k,2k,4k,8k,10k)

12point(100,120,200,
400,800,1k,2k,4k,8k,
10k,15k,20k)

16point(100,120,200,
400,800,1k,2k,4k,8k,
10k,15k,20k,40k,50k,
80k,100k)

Measurement Display Speed2 times/second (slow), 4 times/second (medium speed), 8 times/sec (fast)
Internal Bias0-1500mV adjustable, 1mV stepping
Test Level100mV,300mV,600mV,1000mV,1500mV,2000mV
Correction FunctionOpen circuit correction, short circuit correction
Screening functionThe screening limit can be set from -50% to + 50% with fixed points of 1%, 5%, 10% and 20%
Selection of comparatorFive gear sorting, third gear qualified, one gear failed, one ancillary gear
Communication InterfaceStandard configuration: USB, RS232 (or 485), Handle interface; Optional: GPIB, USB Host
OthersSupport DCR, electrolytic capacitor measurement mode, adjustable backlight brightness, optional in English and Chinese


Quality Bench Type LCR Digital Meters Inductance Measuring Instrument for sale
Send your message to this supplier
 
*From:
*To: XI'AN BEICHENG ELECTRONICS CO.,LTD
*Subject:
*Message:
Characters Remaining: (0/3000)
 
Inquiry Cart 0